30
:TRIGger Commands

:TRIGger:GLITch:LEVel

Command Syntax
:TRIGger:GLITch:LEVel <level_argument>
<level_argument> ::= <level>[, <source>]
<level> ::= .75 x full-scale voltage from center screen in NR3 format
<level> ::= ±(external range setting) in NR3 format
<level> ::= ±8 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger:GLITch:LEVel command sets the trigger level voltage for the
active pulse width trigger.
Query Syntax
:TRIGger:GLITch:LEVel?
The :TRIGger:GLITch:LEVel? query returns the trigger level of the current
pulse width trigger mode. If all channels are off, the query returns
"NONE."
Return Format
<level_argument><NL>
See Also
876
(see
page
1088)
for internal triggers
for external triggers (DSO models)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 851
":TRIGger:GLITch:SOURce"
":EXTernal:RANGe"
on page 304
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 843
on page 880