:POWer:PSRR:FREQuency:MAXimum
Command Syntax
:POWer:PSRR:FREQuency:MAXimum <value>[suffix]
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
[suffix] ::= {Hz | kHz| MHz}
The :POWer:PSRR:FREQuency:MAXimum command sets the end sweep
frequency value. The PSRR measurement is displayed on a log scale, so
you can select from decade values in addition to the maximum frequency
of 20 MHz.
Query Syntax
:POWer:PSRR:FREQuency:MAXimum?
The :POWer:PSRR:FREQuency:MAXimum query returns the maximum
sweep frequency setting.
Return Format
<value><NL>
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
See Also
•
•
•
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
| 20000000}
| 20000000}
":POWer:PSRR:APPLy"
":POWer:PSRR:FREQuency:MINimum"
":POWer:PSRR:RMAXimum"
on page 534
on page 536
on page 537
23
:POWer Commands
535