Trigger:eburst:idle - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

:TRIGger:EBURst:IDLE

Command Syntax
:TRIGger:EBURst:IDLE <time_value>
<time_value> ::= time in seconds in NR3 format
The :TRIGger:EBURst:IDLE command sets the Nth edge in a burst idle
resource in seconds from 10 ns to 10 s. The timer is used to set the
minimum time before the next burst.
Query Syntax
:TRIGger:EBURst:IDLE?
The :TRIGger:EBURst:IDLE? query returns current Nth edge in a burst idle
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :TRIGger Commands"
":TRIGger:EBURst:SLOPe"
":TRIGger:EBURst:COUNt"
on page 843
on page 864
on page 862
30
:TRIGger Commands
863

Advertisement

Table of Contents
loading

Table of Contents