20
:MEASure Power Commands

:MEASure:RIPPle

Command Syntax
:MEASure:RIPPle [<source>]
<source> ::= {CHANnel<n>| FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:RIPPle command installs an output ripple measurement on
screen.
The <source> parameter is the channel probing the output voltage. This
source can also be specified by the :MEASure:SOURce command.
Output ripple is: Vmax - Vmin.
Query Syntax
:MEASure:RIPPle? [<source>]
The :MEASure:RIPPle query returns the measured output ripple.
Return Format
<return_value><NL>
<return_value> ::= the output ripple value in NR3 format