:MEASure:STATistics:INCRement

Command Syntax
:MEASure:STATistics:INCRement
This command updates the statistics once (incrementing the count by one)
using the current measurement values. It corresponds to the front panel
Increment Statistics softkey in the Measurement Statistics Menu. This
command lets you, for example, gather statistics over multiple pulses
captured in a single acquisition. To do this, change the horizontal position
and enter the command for each new pulse that is measured.
This command is only allowed when the oscilloscope is stopped and quick
measurements are on.
The command is allowed in segmented acquisition mode even though the
corresponding front panel softkey is not available.
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :MEASure Commands"
":MEASure:STATistics"
":MEASure:STATistics:DISPlay"
":MEASure:STATistics:RESet"
":MEASure:RESults"
on page 415
on page 388
on page 423
on page 424
on page 427
19
:MEASure Commands
425