30
:TRIGger Commands

:TRIGger:TRANsition:QUALifier

Command Syntax
:TRIGger:TRANsition:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan}
The :TRIGger:TRANsition:QUALifier command specifies whether you are
looking for rise/fall times greater than or less than a certain time value.
The time value is set using the :TRIGger:TRANsition:TIME command.
Query Syntax
:TRIGger:TRANsition:QUALifier?
The :TRIGger:TRANsition:QUALifier? query returns the current rise/fall
time trigger qualifier setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS}
See Also
904
(see
page
1088)
"Introduction to :TRIGger Commands"
":TRIGger:TRANsition:TIME"
":TRIGger:MODE"
on page 851
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 843
on page 907