:MEASure:TSTArt

Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
The short form of this command, TSTA, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:X1Position command (see
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1
cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
page
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:XDELta"
on page 370
":MEASure:TDELta"
on page 1019
":MEASure:TSTOp"
on page 1024
Obsolete and Discontinued Commands
1090). The normal short form "TST" would be the same for
on page 364
on page 388
on page 366
on page 368
34
page
366) instead.
1023