26
:SBUS<n> Commands
:SBUS<n>:A429:TRIGger:PATTern:SSM
Command Syntax
:SBUS<n>:A429:TRIGger:PATTern:SSM <string>
<string> ::= "nn" where n ::= {0 | 1 | X}, length always 2 bits
The :SBUS<n>:A429:TRIGger:PATTern:SSM command defines the ARINC 429
two- bit SSM pattern resource according to the string parameter. This
pattern controls the SSM pattern searched for in each ARINC 429 word.
The specified SSM is only used if the :SBUS<n>:A429:FORMat includes the
SSM field.
Query Syntax
:SBUS<n>:A429:TRIGger:PATTern:SSM?
The :SBUS<n>:A429:TRIGger:PATTern:SSM? query returns the current
settings of the specified ARINC 429 two- bit SSM pattern resource in the
binary string format.
Return Format
<string><NL> in nondecimal format
Errors