:WGEN Commands
:WGEN:OUTPut:LOAD
Command Syntax
:WGEN:OUTPut:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :WGEN:OUTPut:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However,
the output load selection lets the waveform generator display the correct
amplitude and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the
displayed amplitude and offset levels will be incorrect.
Query Syntax
:WGEN:OUTPut:LOAD?
The :WGEN:OUTPut:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also
•
978
(see
page
1088)
"Introduction to :WGEN Commands"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 961