Print Formats - Thermo Scientific 5030 Instruction Manual

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Troubleshooting

Print Formats

Print Formats
6-12
Model 5030 SHARP Monitor Instruction Manual
"status" has 4 figures:
2 figures "Operation Status"
2 figures "Global Error (Break-Down) Status"
("Warning and Limit Status" is not included!)
By Gesytec Protocol Status Version "compatible to FH62I-N":
Operation Status: data bit D1:
"off-line, keyboard enabled"
data bit D2: "filter change"
Global Error (Break-Down) Status:
data bit D0: "air flow rate error"
data bit D1: "filter tape break or filter tape end"
data bit D2: "offset > ± 100%"
data bit D3: "mass buffer is not filled"
In some print formats (for example, no. 1, , no. 30) the instrument status is
provided in addition to the measured value. For example: the device status
in the format "02 00 CA" is given out.
The first two figures represent the "Global Error Status," the two in the
middle represent the "Warning and Limit Statesman the last two represent
the "Operation-Status."
Example: Print format 3:
03-11-16
14:48:18
0200CA 10.3
129.7
1000.7
Thermo Fisher Scientific

Advertisement

Table of Contents
loading

Table of Contents