Dionex ED50A Electrochemical Detector
C.2.9

Signal Statistics

This screen enables you to monitor the selected input to the A/D circuitry.
When the screen is first displayed, the
equal to the
reported in A/D volts and are updated dynamically.
Help Message
Screen Field
CELL
INPUT
MAX
MIN
DURATION
C-30
value, and
INPUT
SIGNAL STATISTICS
CELL:
NORM
INPUT:
C FINE
MAX:
MIN:
DURATION:
Figure C-26. Signal Statistics Screen
Description
Selects the cell test mode:
NORM
Uses input data from the cell.
FIXED
Uses a fixed dummy cell for input.
The measured data from the A/D circuit:
C FINE
Conductivity cell fine input
C COARSE
Conductivity cell coarse input
C THER
Conductivity cell thermistor input
E FINE
Amperometry cell fine input
E COARSE
Amperometry cell coarse input
E pH
Amperometry cell pH input
LEAK
Leak detector input
DRIVE
Cell drive circuit input
The maximum input data value during the duration of the test.
The minimum input data value during the duration of the test.
The duration (in minutes) of the test. The test starts when you enter
this screen and terminates when you exit.
and
MAX
reads 0. Status values are
DURATION
n.nnn volts
n.nnn volts
nnn.nn Minutes
status values are
MIN