Maintenance

8.4.10 Cleaning the Densitometer Calibration Plate

Purpose:
To ensure optimum calibration of the densitometer.
Special tools
– Damp lint-free cloth
– Soft brush
CAUTION!
Risk of injuries by moving transport rollers. Open the cover of the paper outlet
(PO) only after the transport rollers have stopped or when the machine is switched
off.
2
1
G_GG115
2
dlab2515
8-36
2005-05-01 / PN 04092_01
1
d-lab.2, d-lab.2plus, netlab.2plus: Operation
Remove the front cover of the paper outlet
PO (1): loosen the screw (2).
Applies to d-lab.2
Unscrew the knurled screws (1) on the right
and left side on the guide plate of the
densitometer unit.
Fold up the densitometer unit (2) and clean
the calibration plate with a damp cloth.
Clean the guide plates with a damp lint-free
cloth.
Clean the densitometer head with a brush.
Fasten the densitometer unit again with the
knurled screws.
Fasten the paper outlet cover.
Note
Damaged calibration plates must be replaced
by a service technician.