Changing Chemical Solutions (Msc/D-Lab Chemicals) - AgfaPhoto d-lab.2 8060/260 Operation Manual

Table of Contents

Advertisement

Changing Chemicals
3.3

Changing Chemical Solutions (MSC/d-lab Chemicals)

G_GG201_3 (Fig. shows d-lab.2)
Note
Mixing the MSC/d-lab chemicals for first machine operation is explained in the
Installation folder. If the machine is put back in operation after a longer down time
(more than two weeks, see
Installation folder.
CAUTION!
Whenever you work with chemicals observe the
beginning of this chapter!
The chemical tank solutions (developer – CD, bleachfix – BX, stabilizer – SB) are changed
in several steps:
1. Prepare the machine tanks (drain the old chemicals, clean the tanks, and check the
hoses).
2. Remove the chemical filters.
3. Mix the MSC/d-lab chemicals.
4. Insert the new chemical filters.
Afterwards
the replenisher chemicals must be mixed (see
(d-lab.2 Easy Paper Box))
various settings must be entered or checked
Settings and Tests for Chemicals
(see
tests must be made (see
3-4
2005-05-01 / PN 04092_01
1
6
5
4
3
2
Chapter 11) mix the chemicals also as described in the
in this chapter);
in this chapter);
Settings and Tests for Chemicals
d-lab.2, d-lab.2plus, netlab.2plus: Operation
1 Wet section
2 LED docking station
3 Drain taps for machine tanks
4 Docking station
5 Replenisher tanks
6 Water tanks
Safety Information
at the
Changing Replenisher Chemicals
in this chapter).