Programming an Adjacent Channel Leakage Ratio Measurement
• Because multi-measurements are being used, Low_1st_alt and Hi_1st_alt (± 5 MHz) measurements
return the average adjacent channel leakage ratio measurement values for those offsets, in dB. If
multi-measurements were not being used, the returned values would be a single measurement value for
each frequency offset.
• Because the Low_2nd_alt and Hi_2nd_alt (± 10 MHz) measurements are turned off in this example, the
returned values are always 9.91E+37 (NAN).
Related Topics
•
"Adjacent Channel Leakage Ratio Troubleshooting"
•
"Adjacent Channel Leakage Ratio Measurement Description"
•
"SETup:WACLeakage"
•
"INITiate"
•
"FETCh:WACLeakage"
•
"Amplitude Offset"
•
"Integrity Indicator"
136