FETCh:AFANalyzer
FETCh:AFANalyzer:SINAD[:AVERage]?
Function
Queries the average SINAD measurement.
Query
Average SINAD
• Range:-99 to 99 dB, 9.91 E+37
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:AFANALYZER:SINAD:AVERAGE?"
FETCh:AFANalyzer:SINAD:ALL?
Function
Queries the SINAD measurements.
Measurement results are returned in the following order: minimim SINAD, maximum SINAD,
average SINAD, and standard deviation SINAD.
Query
Minimum, maximum, average SINAD:
• Range: -99 to 99 dB, 9.91 E+37
• Resolution: 0.01 dB
Standard deviation SINAD
• Range: -99 to 99 dB, 9.91 E+37
• Resolution: 0.001 dB
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:AFANALYZER:SINAD:ALL?"
!Returns mininum, maximum, average, and standard deviation SINAD.
FETCh:AFANalyzer:SINAD:MAXimum?
Function
Queries the maximum SINAD measurements.
Query
Range: -99 to 99 dB, 9.91 E+37
Resolution: 0.01 dB
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:AFANALYZER:SINAD:MAXIMUM?"