Agilent Technologies 8960 Series Reference Manual page 109

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

The test set's operating mode is set using the following command:
Command to Set the Test Set's Operating Mode
Command
CALL:OPER:MODE <operating mode>
• Set Amplitude Offsets
To achieve accurate measurement results, it is important to account for losses in the cabling and fixturing
between the mobile station and test set. You must determine what the losses are for your test setup and
then specify the appropriate frequency-dependent amplitude offset values.
You can specify amplitude offsets for up to 20 frequencies using the following commands:
Commands to Set Amplitude Offsets
Command
SYST:CORR:FREQ <freq1>, <freq2>,
<freq3>
SYST:CORR[:SGAin] <offset1>,
<offset2>, <offset3>
• Set Display Mode
To achieve a slightly faster test execution speed, you can disable the front panel display on the test set
using the DISP:MODE FAST command.
Example
CALL:OPER:MODE CELL
Example
SYST:CORR:FREQ 800MHz, 1800MHz,
1900MHz
SYST:CORR -0.7, -1.0, -1.2
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_prog_generic_step1.fm
Step 1: Set Up the Test Set
109

Advertisement

Table of Contents
loading

Table of Contents