Agilent Technologies 8960 Series Reference Manual page 334

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

FETCh:WCDomain
FETCh:WCDomain:FERRor?
Function
Queries the worst case frequency error.
Query
Range: -99 to 99 kHz, 9.91 E+37
Resolution: 0.1
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WCDOMAIN:FERROR?" !Returns worst case frequency error.
FETCh:WCDomain:INTegrity?
Function
Queries the measurement integrity indicator for the last code domain measurement completed.
Zero indicates a normal measurement. See
about non-zero integrity results.
Query
Range: 0 through 23
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WCDOMAIN:INTEGRITY?"
!Queries the integrity indicator for the code domain measurement results.
FETCh:WCDomain:MERRor?
Function
Queries the maximum magnitude error.
Query
Range: 0.0 to 99.0%
Resolution: 0.1
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WCDOMAIN:MERRor?" !Returns maximum magnitude error.
FETCh:WCDomain:OOFFset?
Function
Queries the maximum origin offset.
Query
Range: -99 to 0 dB, 9.91 E+37
Resolution: 0.1
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WCDOMAIN:OOFFSET?" !Returns maximum origin offset.
334
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_fetch_wcdomain.fm
"Integrity Indicator" on page 190
for information

Advertisement

Table of Contents
loading

Table of Contents