• Code Domain Power (CDP)
• Code Domain Error (CDE)
• All of the measurement results of the Waveform Quality measurement.
Measurement Description" on page 44.
Input Signal Requirements
• This measurement does not autorange, therefore the expected power level from the UE is set using the
current MS Target Power setting, or by using manual receiver power control and specifying the manual
power level. See
"Receiver Control"
The frequency of the signal being measured must be in the range of 800 MHz to 1000 MHz, or from 1700 MHz to
•
1990 MHz.
The level into the test set's RF IN/OUT connector must be in the range of +5 dBm to +28 dBm, in a 3.84 MHz
•
bandwidth.
• Maximum measurable EVM = 35%
• Maximum measurable frequency error = ± 1 kHz
.
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_meas_cod_dom_desc.fm
Code Domain Measurement Description
See "Waveform Quality
51