SYSTem:COMMunicate:GPIB:DEBug[:STATe]
Function
Sets/queries the test set's SCPI debugger state.
When the state is on; enhanced error messages (generated from GPIB commands with syntax
errors) are shown the test set display.
The error message is printed along with the syntax. <ERR> is displayed at the end of the
incorrect node. Non-printable characters will be replaced with the $ symbol. See
Messages" on page 588
The debugger state should be set to on only during GPIB code development. Test times will
increase if the debugger state is left on.
Setting
Range: 0|OFF | 1|ON
Query
0|1
*RST setting
0 (off)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SYSTEM:COMMUNICATE:GPIB:DEBUG:STATE ON" !Sets debugger to on.
for a list of the errors.
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_system_communicate.fm