CALL:CCPChannel:PRIMary[:SLEVel]
Function
Sets/queries the initial power offset of the primary common control physical channel . This
command sets the power offset level and turns on the power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this value.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-9.9 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CCPCHANNEL:PRIMARY:SLEVEL -15DB" !Enables and sets the initial P-CCPCH power
offset to -15 dB.
CALL:CCPChannel:PRIMary:LEVel
Function
Sets/queries the primary common control physical channel's initial power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this setting.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-9.9 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CCPCHANNEL:PRIMARY:LEVEL -15DB" !Sets the initial P-CCPCH power offset to -15 dB.
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_call_ccpchannel.fm
CALL:CCPChannel
231