INITiate
INITiate:<measurement mnemonic>:OFF
Function
Deactivates the selected measurement. See
Only one measurement can be deactivated at a time. To stop one or more measurements and
leave them in the active state, use
Requirements
Test Application Revision: A.01 and above
Test Application Revision: A.02 added WCDomain, WSEMask, and WOBWidth.
Programming Example
OUTPUT 714;"INITIATE:WCPOWER:OFF" !Deactivates the channel power measurement.
INITiate:COUNt?
Function
Queries the number of measurements that have been initiated (are active). See
States" on page
Query
Range: 0 to 10
Requirements
Test Application Revision: A.01 and above
Test Application Revision: A.02 added WCDomain, WSEMask, and WOBWidth.
Programming Example
OUTPUT 714;"INITIATE:COUNT?"
INITiate:DONE?
Function
Queries (one at a time) which measurements, if any, are finished or have timed-out.
See
"Measurement Event Synchronization" on page 201
See
"Measurement States" on page 187
See
"INITiate:DONE:FLAG:<measurement mnemonic>" on page 379
commands.
Requirements
Test Application Revision: A.01 and above
Test Application Revision: A.02 added WCDomain, WSEMask, and WOBWidth.
Query
Range: AFAN|WACL|WBER|WCD|WCP|WOBW|WSEM|WTP|WWQ|NONE|WAIT
INITiate:DONE:CLEar
Function
Clears the done flag from all measurements.
Requirements
Test Application Revision: A.01 and above
Test Application Revision: A.02 added WCDomain, WSEMask, and WOBWidth.
Programming Example
OUTPUT 714;"INITIATE:DONE:CLEAR" !Clears the done flag.
378
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_initiate.fm
"ABORt" on page
187.
to understand the test set's measurement states.
"Measurement States" on page
213.
for how to use this command.