Agilent Technologies 8960 Series Reference Manual page 254

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

CALL:CPIChannel
CALL:CPIChannel:LEVel
Function
Sets/queries the common pilot channel's initial power offset. This command sets the power offset
level and turns on the power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this value.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-4.8 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CPICHANNEL:LEVEL -5"
CALL:CPIChannel:STATe
Function
Sets/queries the common pilot channel's state.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this value.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
1 (on)
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CPICHANNEL:STATE ON"
254
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_call_cpichannel.fm

Advertisement

Table of Contents
loading

Table of Contents