Agilent Technologies 8960 Series Reference Manual page 419

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:WCDomain:CONTinuous
Function
Sets/queries the trigger arm state for the code domain measurement. The trigger arm state
determines whether the bit error measurement will be made once then stop (0|OFF (single)), or
automatically re-arm upon completion of one measurement and repeat the process (1|ON
(continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCDOMAIN:CONTINUOUS ON
!Sets measurement to continuous trigger mode.
SETup:WCDomain:TIMeout[:STIMe]
Function
The setting form of this command performs two functions: (1) sets the code domain
measurement's timeout (see
leakage measurement's timeout STATe to 1 (on).
The query form returns the loopback bit error rate measurement's timeout value.
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
20 seconds, on
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCDOMAIN:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
SETup:WCDomain:TIMeout:STATe
Function
Sets/queries the code domain measurement's timeout state (see
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCDOMAIN:TIMEOUT:STATE ON"
"
"Measurement Timeouts"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_setup_wcdomain.fm
) value, and (2) sets the adjacent channel
"Measurement Timeouts"
SETup:WCDomain
).
419