Agilent Technologies 8960 Series Reference Manual page 433

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:WSEMask:TIMeout:TIME
Function
Sets/queries the spectrum emission mask measurement's timeout (see
setting to be used when the timeout state is set to ON.
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
10 seconds
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:TIME 5 S"
SETup:WSEMask:TRIGger:DELay
Function
Sets/queries the spectrum emission mask measurement's trigger delay setting.
This setting can be used to delay the measurement trigger in order to to start a measurement
during a specific timeslot in the frame. For example, a 1.333 ms delay would cause the trigger to
wait until after the first two timeslots have elapsed.
Setting
Range: -10 to 10 milliseconds
Resolution: 1 nanosecond
Query
Range: -10 to 10 milliseconds
*RST Setting
0 seconds
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TRIGGER:DELAY 1MS"
SETup:WSEMask:TRIGger:SOURce
Function
Sets/queries the spectrum emission mask measurement's trigger source.
Setting
Range: IMMediate|PROTocol|AUTO
Query
Range: IMM|PROT|AUTO
*RST Setting
AUTO
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TRIGGER:SOURCE IMMEDIATE"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_setup_wsemask.fm
SETup:WSEMask
"Measurement Timeouts"
)
433

Advertisement

Table of Contents
loading

Table of Contents