READ:WSEMask[:ALL]?
Function
Initiates and fetches a spectrum emissions mask measurement as a sequential operation.
Measurement results are returned in the following order: integrity, overall pass/fail, in-channel
power, and pass/fail results for each of the four frequency ranges.
In-channel power is returned in dBm/3.84 MHz.
The four frequency ranges for this measurement are as follows:
• Range 1: +/- 2.515 to 3.485 MHz
• Range 2: +/- 4.000 to 7.500 MHz
• Range 3: +/- 7.500 to 8.500 MHz
• Range 4: +/- 8.500 to 12.500 MHz
If the multi-measurement state is on, the result represents the average of a number of spectrum
emissions measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
Integrity
• Range: 0 to 23
Overall Pass/Fail
• Range: 0|1
In-Channel Power (dBm/3.84 MHz)
• Range: -99.00 to +99.00 dBm, 9.91 E+37
• Resolution: 0.01
Range Pass/Fail
• Range: 0|1
Requirements
Test Application Revision: A.02 and above
READ:WSEMask:FAIL?
Function
Initiates and fetches a spectrum emissions mask measurement as a sequential operation.
Measurement results are returned in the following order: integrity, overall pass/fail (0 = Pass, 1 =
Fail).
If the multi-measurement state is on, the result represents the average of a number of thermal
power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
• Range: 0|1
Requirements
Test Application Revision: A.02 and above
"SETup:WSEMask:COUNt[:SNUMber]"
"SETup:WSEMask:COUNt[:SNUMber]"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_read.fm
.
.
READ
387