Use the following procedure to manually test an 8960 that is running an
E1960A GSM Test Application. A test record is included to assist in
collecting test data. Make a copy of the test record to use.
The test equipment required for 8960 Operational Verification is:
ESG E4433B Digital Signal Generator
VSA E4406A Transmitter Tester
8960
RF In/Out
To perform the 8960 Operational Verification tests requires that the 10 MHz
timebases be connected together. Connect the 10 MHz Reference Output of the 8960
to the 10 MHz Ref In of the ESG E4433B. Connect the 10 MHz Ref Out of the ESG
E4433B to the 10 MHz Ref In of the VSA E4406A with BNC cables.
The RF In/Out Connector of the 8960 will be alternately connected to either the RF
Out of the ESG E4433B (RX Test) or the RF Input of the VSA E4406A (TX Test) with
a low loss Type-N cable.
Agilent Technologies 8960
Operational Verification (Manual
Procedure)
Test Equipment Setup
10 MHz
Ref Out
10 MHz
Ref In
ESG
E4433B
Audio In
LF Out
Hi
10 MHz
Ref Out
VSA
E4406A
RF Out
(TX Test)
(RX Test)
Manual
Verification
10 MHz
Ref In
RF In
1

Advertisement

loading

Summary of Contents for Agilent Technologies 8960

  • Page 1 E4433B to the 10 MHz Ref In of the VSA E4406A with BNC cables. The RF In/Out Connector of the 8960 will be alternately connected to either the RF Out of the ESG E4433B (RX Test) or the RF Input of the VSA E4406A (TX Test) with...
  • Page 2 Manual 8960 Operational Verification Verification Test Record Test Data Record for 8960 with E1960A GSM Test Application Functional Verification Expected Limit (Note 1) Pass Fail ± 2.0 dB Analog Generator Level Accy Analog Generator Spectral Purity ≤ -25 dBc Harmonics ≤...
  • Page 3 Verification 8960 Operational Verification Analog Generator Level Accuracy Analog Generator Level Accuracy is a test to insure that the 8960 can set a basic level and then step in 10 dB increments accurately. The expected limit is: RF Generator Output Level, RF In/Out (2 sources) - ±...
  • Page 4 Verification 8960 Operational Verification Analog Generator Level Accuracy, cont. On the 8960 perform the following set up functions: 1. Press the blue ‘ SHIFT’ button and the ‘ PRESET’ button (front panel buttons not shown above). 2. Press the display ‘ Operating Mode’ button and set mode to Test using the knob.
  • Page 5 Manual Verification 8960 Operational Verification Analog Generator Level Accuracy, cont. On the E4406A, press the ‘ Preset’ System button, the ‘ MODE’ Control button and set to ‘ Basic’ Mode . On the E4406A, press the ‘ FREQUENCY’ Control button and set the...
  • Page 6 Manual Verification 8960 Operational Verification Analog Generator Level Accuracy, cont. On the E4406A, press the ‘ ZOOM’ button, the ‘ Marker’ button, and the ‘ Search’ button. Select the ‘ TRACE’ screen button and set the trace to Spectrum Average.
  • Page 7 8960 Operational Verification Analog Generator Level Accuracy, cont. Step the 8960 down in amplitude in 10 dB increments to -80 dBm. Insure that each level step is accurate within ± 2 dB. On the 8960 perform the following set up functions: 1.
  • Page 8 8960 Operational Verification Analog Generator Level Accuracy, cont. On the E4406A set Center Frequency to 1.8052 GHz. Step the 8960 down in amplitude in 10 dB increments to -80 dBm. Insure that each level step is accurate within ± 2 dB.
  • Page 9 The 8960 contains 2 complete signal generator sources (these are very similar to the E4432/33B Signal Generator). The operating firmware of the 8960 only provides for operation of 1 source using the manual user interface (front panel) in the analog mode.
  • Page 10 Verification Analog Generator Spectral Purity Analog Generator Spectral Purity is a test to insure that the 8960 has harmonics and spurious signals within specification.The test is run by setting the 8960 to a carrier frequency of 300 MHz and power level of -10 dBm. Harmonics and spurious are checked at 450MHz, 600 MHz, 750 MHz, and 900 MHz.
  • Page 11 ON the E4406A make the following settings: Center Freq: 300 MHz Span: 10 Press the ‘ ZOOM’ button, the ‘ Marker’ button, and the ‘ Search’ button. Select the ‘ TRACE’ screen button and set the trace to Spectrum Average. The 8960 amplitude should equal -10 dBm ± 2 dB.
  • Page 12 Manual 8960 Operational Verification Verification Analog Generator Spectral Purity On the E4406A set the center frequency to 450 MHz. Press the ‘ AMPLITUDE’ button and set Ref Value to -50 dBm. The Sub-Harmonic expected limit is: ≤ -50 dBm. On the E4406A set Center Frequency to 600 MHz. The Harmonic expected...
  • Page 13 Manual 8960 Operational Verification Verification Analog Generator Spectral Purity On the E4406A set the Center Frequency to 750 MHz. The Sub-Harmonic expected limit is: ≤ -50 dBm On the E4406A set the Center Frequency to 900 MHz. The Harmonic expected limit is: ≤ -35 dBm .
  • Page 14 The Audio Generator test checks to insure that the 8960 has an audio source signal with reasonable performance of accuracy. The test is run by setting the 8960 in the audio generator mode and looping an audio signal from the Audio Out BNC connector to the Audio In Hi BNC connector to measure level accuracy of the audio output.
  • Page 15 Manual Verification 8960 Operational Verification Audio Generator Accuracy,cont. Press the Instrument Select button. Press the knob and activate the Audio Generator screen Set audio amplitude to 1.414 V. Set audio frequency to 1 kHz.
  • Page 16 Manual Verification 8960 Operational Verification Audio Generator Accuracy,cont. Press the ‘ Measurement Selection’ button. Use the knob and select Analog Audio, push the knob to start selection. The Analog Audio measurement screen should appear. The expected limits are: Level Accuracy: ± 0.03V...
  • Page 17 RMS Phase Error, and Frequency Error Source 1 GSM Generator tests check to insure that the 8960 has a GSM Digital signal within a reasonable specification.The test is run by setting the 8960 to a traffic channel of 30 and power level of -10 dBm.
  • Page 18 Manual Verification 8960 Operational Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 1 cont. Go to Screen 2 of 3. Set the Traffic Channel to 124 . Go to Screen 3 of 3. Set the expected burst to TSC0.
  • Page 19 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 1 cont. On the E4406A, press the ‘ MODE’ Control button and set Mode field to GSM. On the E4406A set the Center Frequency to 959.8 MHz (channel 124). The...
  • Page 20 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 1 cont. On the E4406A, press the PVTbutton. On the E4406A, press the ‘ AMPLITUDE’ Control button. Set ‘ Scale/Div’ field to be .3 dB and the ‘...
  • Page 21 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 1 cont. On the E4406A, press the ‘ Phase & Freq’ button. The display shown above should appear. The expected limits for Phase & Frequency are: Peak Phase Error <±...
  • Page 22 RMS Phase Error, and Frequency Error Source 2 GSM Generator tests check to insure that the 8960 has a GSM Digital signal within a reasonable specification.The test is run by setting the 8960 to a traffic channel of 30 and power level of -10 dBm.
  • Page 23 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 2 cont. Go to Screen 2 of 3. Set the Traffic Band to PGSM and the Traffic Channel to 124. Go to Screen 3 of 3. Set the expected burst to TSC0.
  • Page 24 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 2 cont. On the E4406A, press the ‘ MODE’ Control button and set Mode field to GSM. On the E4406A set the Center Frequency to 959.8 MHz (channel 124). The...
  • Page 25 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 2 cont. On the E4406A, press the PVTbutton. On the E4406A, press the ‘ AMPLITUDE’ Control button. Set ‘ Scale/Div’ field to be .3 dB and the ‘...
  • Page 26 Manual 8960 Operational Verification Verification GSM Generator Amplitude Flatness, Peak Phase Error, RMS Phase Error, and Frequency Error, Source 2 cont. On the E4406A, press the ‘ Phase & Freq’ button. The display shown above should appear. The expected limits for Phase & Frequency are: Peak Phase Error <±...
  • Page 27 8960 Operational Verification Analog Audio Analyzer The Analog Audio Analyzer is tested to insure that the 8960 can accurately measure an audio signal within a reasonable limit. The test is run by setting the 8960 into the Audio Analyzer measurement mode.
  • Page 28 Manual Verification 8960 Operational Verification Analog Audio Analyzer, cont. Press the ‘ Measurement Selection’ button. Use the knob and select Analog Audio, push the knob to start selection. LF Out FREQUENCY AMPLITUDE -135.0 4.000 000 000 00 LF Out Amplitude 1.414V...
  • Page 29 Manual Verification 8960 Operational Verification Analog Audio Analyzer, cont. The 8960 Analog Audio screen should display a voltage level measurement of the audio signal from the E4433B . The Analog Audio measurement expected limit is ± 0.04V...
  • Page 30 8960 Operational Verification Analog Analyzer RF Power Meter The Analog Analyzer RF Power Meter is tested to insure that the 8960 can accurately measure an RF signal within a reasonable limit. The test is run by setting the 8960 into a manual measurement mode. The E4433B is then set to output a CW RF signal at various levels and frequencies.
  • Page 31 8960 Operational Verification Verification Analog Analyzer RF Power Meter, cont. On the 8960 perform the following set up functions: 1. Press the blue ‘ SHIFT’ button and the ‘ PRESET’ button (front panel buttons not shown above). 2. Press the display ‘ Operating Mode’ button and set mode to Test using the knob.
  • Page 32 Manual Verification 8960 Operational Verification Analog Analyzer RF Power Meter, cont. Press ‘ Transmit Power Setup’ button and set Trigger Source to ‘ Immediate’ . On screen 2 of 2 set Expected Power to +13.00 dBm. RF Power measurement Specification is < ± 1.63 dB.
  • Page 33 Analog Analyzer RF Power Meter, cont. FREQUENCY AMPLITUDE 939.000 000 00 -15.00 EXT REF On the E4433B set the Amplitude to -15 dBm. On the 8960 set Expected Power to -15.00 dBm. RF Power measurement expected limit is < ± 1.60 dB.
  • Page 34 13.00 EXT REF Change the E4433B amplitude to +13 dBm and frequency to 1.85 GHz . On screen 2 of 2 of the 8960 set expected power to +13.00 dBm. RF Power measurement expected limit is < ± 1.60 dB.
  • Page 35 Verification 8960 Operational Verification GSM Analyzer The GSM Analyzer is tested to insure that the 8960 can accurately measure a GSM burst signal within a reasonable limit. The test is run by setting the 8960 into GSM Analyzer manual measurement mode.
  • Page 36 8960 Operational Verification Verification GSM Analyzer On the 8960 perform the following set up functions: 1. Press the blue ‘ SHIFT’ button and the ‘ PRESET’ button (front panel buttons not shown above). 2. Press the display ‘ Operating Mode’ button and set mode to Test using the knob.
  • Page 37 Manual 8960 Operational Verification Verification GSM Analyzer, cont. Press ‘ Measurement Selection’ button and select ‘ Transmit Power’ as the Operating Mode. Transmit Power window should display a reading of approximately +13 dBm Press ‘ Measurement Selection’ button and select ‘ Phase & Frequency Error’ as the Operating Mode.
  • Page 38 Manual 8960 Operational Verification Verification GSM Analyzer, cont. The Phase and Frequency Error window should appear and display average readings of Peak and RMSPphase error and Frequency error. The expected measurement limits are: Peak error is <± 8 Deg, RMS error is <±2 Deg, Frequency error is <±24 Press ‘...
  • Page 39 Manual 8960 Operational Verification Verification GSM Analyzer, cont. Press ‘ Return to PvT Control’ button then press the ‘ Change View’ button. Press the ‘ Numeric 1 of 2’ button. The Power vs Time Numeric Screen 1 should appear. The expected limits are: ≤...
  • Page 40 -14 dBc at a 552.8 usec offset on the burst. Variation in burst timing can vary the offset in dBc greatly. This measurement is meant to show consistency between 8960 units using the same identical source for each unit tested.
  • Page 41 Manual Verification 8960 Operational Verification GSM Analyzer, cont. The ORFS measurement screen must be configured for offset frequency values to make a measurement. Press the ‘ Modulation Frequencies’ button. Load the default frequencies into each offset by rotating the knob to each offset and pressing the ‘...
  • Page 42 ORFS performance of -79 dB at 600 kHz from the carrier. The values may change however according to the source. The measured accuracy of a source should be consistent for every 8960 unit. The measurement done for this test is meant to show test consistency between 8960 units using the same identical source...