Agilent Technologies 8960 Series Reference Manual page 439

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:WWQuality:CONTinuous
Function
Sets/queries the trigger arm state for the waveform quality measurement. The trigger arm state
determines whether the waveform quality measurement will be made once then stop (0|OFF
(single)), or automatically rearm upon completion of one measurement and repeat the process
(1|ON (continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WWQUALITY:CONTINUOUS OFF
!Sets measurement to single trigger mode.
SETup:WWQuality:COUNt[:SNUMber]
Function
The setting form of this command performs two functions: (1) sets the number of waveform
quality measurements to be made in a multi-measurement, and (2) sets the multi-measurement
STATe command to 1 (on).
The query form returns the number of waveform quality multi-measurements to be made when
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
Off (10 when STATe = On)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WWQUALITY:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
SETup:WWQuality:COUNt:NUMBer
Function
Sets/queries the number of waveform quality measurements to be made when the
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
10
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WWQUALITY:COUNT:NUMBER 5"
!Sets the number of measurements to 5 when the multi-measurement state is set to ON.
"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_setup_wwquality.fm
SETup:WWQuality
439