SETup:WBERror:TIMeout:STATe
Function
Sets/queries the loopback bit error ratio measurement's timeout state (see
Timeouts"
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WBERROR:TIMEOUT:STATE ON"
SETup:WBERror:TIMeout:TIME
Function
Sets/queries loopback bit error ratio measurement's timeout setting to be used when the timeout
state is set to ON (see
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
20 seconds
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WBERROR:TIMEOUT:TIME 5 S"
).