Agilent Technologies 8960 Series Reference Manual page 487

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

STATus:QUEStionable:ERRors:CDMA Register Bit Assignments
The STATus:QUEStionable:ERRors:CDMA register bits will be used to indicate information about CDMA
related test set device-specific errors (positive error numbers).
Bit
Binary
Number
Weighting
15
32768
14
16384
13
8192
12
4096
11
2048
10
1024
9
512
8
256
7
128
6
64
5
32
Condition
Not Used. Defined by SCPI.
MUI Maskable Message.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
+900 Errors
+800 Errors
+700 Errors
+600 Errors
+500 Errors
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_status_questionable.fm
Description
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +900 to +999 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +800 to +899 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
The condition bit will be pulsed to a 1 and
immediately back to 0 if an error in the +700 to
+799 range occurs. After setting the Positive
Transition Filter and the Negative Transition Filter
you may query the Event Register to find out which
of these errors occurred.
The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +600 to +699 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +500 to +599 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
STATus:QUEStionable
487