FETCh:WSEMask
FETCh:WSEMask:ICOunt?
Function
Queries the intermediate count of measurements completed in a multi-measurement.
To set up multi-measurements, see
Query
Range: 0 to 999
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WSEMASK:ICOUNT?"
!Returns the intermediate count of spectrum emissions mask multi-measurements.
FETCh:WSEMask:ICPower?
Function
Queries the in-channel absolute power. To set the number of measurement to average, use the
command
Query
• Range: -99 to 99 dBm, 9.91 E+37
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WSEMASK:ICPOWER?"
!Queries the in-channel power for the spectrum emissions mask results.
FETCh:WSEMask:INTegrity?
Function
Queries the measurement integrity indicator for the last spectrum emissions mask measurement
completed.
Zero indicates a normal measurement. See
about non-zero integrity results.
Query
Range: 0 through 23
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"FETCH:WSEMASK:INTegrity?"
!Queries the integrity indicator for the channel spectrum emissions mask results.
348
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_fetch_wsemask.fm
"SETup:WSEMask:COUNt[:SNUMber]" on page
"SETup:WSEMask:COUNt[:SNUMber]"
431.
"Integrity Indicator" on page 190
.
for information