SETup:WSEMask
SETup:WSEMask:COUNt:STATe
Function
Sets/queries the spectrum emission mask multi-measurement state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:COUNT:STATE ON" !Turns on spectrum emission mask
multi-measurements.
SETup:WSEMask:TIMeout[:STIMe]
Function
The setting form of this command performs two functions: (1) sets the spectrum emission mask
measurement's timeout (see
mask measurement's timeout STATe to 1 (on).
The query form returns the spectrum emission mask measurement's timeout value.
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
10 seconds, on
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
SETup:WSEMask:TIMeout:STATe
Function
Sets/queries the spectrum emission mask measurement's timeout (see
state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:STATE ON"