For example, the transmitter power measurement results can be returned as:
Example FETCh? Result Formats
Command
FETC:TXP?
FETC:TXP:POW:ALL?
FETC:TXP:POW:AVER?
FETC:TXP:POW:MIN?
FETC:TXP:POW:MAX?
FETC:TXP:POW:SDEV?
FETC:TXP:INT?
Example 12. Concurrent Measurement Process Programming Example
! Start a Set of Concurrent Measurements:
!
OUTPUT Test_set;"INIT:TXP;PFER"
!
! Determine if a Measurement Is Done:
!
LOOP
OUTPUT Test_set;"INIT:DONE?"
ENTER Test_set;Meas_done$
!
! Obtain a Set of Measurement Results:
!
SELECT Meas_done$
CASE "TXP"
OUTPUT Test_set;"FETC:TXP:POW?"
ENTER Test_set;Avg_tx_power
CASE "PFER"
OUTPUT Test_set;"FETC:PFER:RMS?"
ENTER Test_set;Max_rms_phas_er
END SELECT
EXIT IF Meas_done$="NONE"
END LOOP
Results Returned
Measurement integrity and average value
Minimum, maximum, average and standard
deviation values
Average value only
Minimum value only
Maximum value only
Standard deviation value only
Measurement integrity value only
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_prog_generic_step5.fm
Step 5: INITiate and FETCh Measurements