SETup:WCPower:COUNt:STATe
Function
Sets/queries the channel power multi-measurement state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WCPOWER:COUNT:STATE ON" !Turns on channel power multi-measurements.
SETup:WCPower:FILTer[:RRC]
Function
Sets/queries the state of the root-raised cosine (RRC) filter.
Setting
Range: ON|OFF
Query
Range: ON|OFF
*RST Setting
OFF
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCPOWER:FILTER:RRC ON" !Turns on the RRC filter.
SETup:WCPower:INTerval:TIME
Function
Sets/queries the channel power measurement interval. 10 milliseconds represents a full frame.
Setting
Range: 0.1 to 10 milliseconds
Resolution: 100 nanoseconds (minimum)
Query
Range: 0.01 to 10 milliseconds
*RST Setting
0.6667 ms (one slot interval)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"SETUP:WCPOWER:INTERVAL:TIME 10MS
!Sets interval time to 10 milliseconds.
"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_setup_wcpower.fm
SETup:WCPower