Measurement Event Synchronization
• Adjacent Channel Leakage Ratio
• Channel Power
• Code Domain
• Loopback Bit Error Ratio
• Occupied Bandwidth
• Spectrum Emission Mask
• Thermal Power
• Waveform Quality
Operating Considerations
Only one indicator is returned per query.
All active measurements must be set to single trigger mode. This ensures that when a measurement completes
it remains in the "DONE" state rather than restarting. Sending the "*RST" command at the beginning of the
test code or using the "SETup:CONTinous:OFF" command during measurement setups are ways to set the
trigger to single for all measurements.
Related Topics
"INITiate Command Functions" on page 374
"What Happens When a Measurement is INITiated?" on page 374
"Concurrent Measurements" on page 180
"STATus:OPERation:NMRReady:<WCDMA | FDD> Register Bit Assignments" on page 466