CALibration Subsystem
Table 5. Measurement Calibration Indicator values
Measurement
Calibration
Indicator
0
Measurement's calibration completed successfully.
1
Measurement's calibration failed due to temperature drift. Wait for temperature to stabilize.
2
Measurement's calibration failed due to oven out of range.
3
Measurement's calibration failed due to loopback switch problem.
4
Measurement's calibration failed due to under range condition.
5
Measurement's calibration failed due to over range condition.
6
Measurement's calibration failed due to correlation problem.
-340
Measurement's calibration failed. This is a SCPI defined value indicating an unexpected case.
You can also manually initiate the measurement calibration for individual measurements by selecting the
measurement and pressing the Calibrate Measurements softkey. This calibration is not an option for the
Thermal Power measurement or any receiver measurements.

Syntax Diagram and Command Descriptions

"CALibration"
222
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_calibration_description.fm
Measurement Calibration Message