Rear Panel Connectors
TRIG IN
This BNC connector provides the capability for you to trigger measurements or instruments from an external
source. Setting the Trigger Source setting to External for measurements or instruments which support this
feature causes the measurement or instrument to trigger when a positive-going TTL compatible pulse is
applied to this connector. Currently, the only measurement or instrument which supports this feature is the
Spectrum Monitor.
TRIG OUT
This BNC connector allows you to synchronize the test set to other equipment and is configured by setting
Frame Trigger Parameters.
VGA OUT
This DB-15 connector allows you to simultaneously route the test set's display to another monitor.
10 MHZ REF IN
This BNC connector accepts an external 10 MHz timebase signal. The nominal input impedance is 50 ohm.
This version of test set can only accept a 10 MHz timebase signal.
10 MHZ REF OUT
This BNC connector provides a 10 MHz timebase signal to external test equipment. The accuracy of this signal
is determined by the timebase used. The nominal output impedance is 50 ohm with a typical level of 0.5 V rms.