CALL[:CELL]:CLPControl:UPLink:MODE
Function
Sets/queries the uplink power control setting when using the Active Cell operating mode.
Setting
Range:
• ACTive (activates closed loop power control)
• UDOWn (alternating bits)
• UP (all up bits)
• DOWN (all down bits)
• UDOWn10 (ten up/ten down bits)
Query
Range: Same as setting range.
*RST Setting
ACTive
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"CALL:CELL:CLPControl:UPLink:MODE ACTive" !Sets the uplink closed loop power control to
active.
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_call_clpcontrol.fm
CALL[:CELL]:CLPControl:UPLink
237