FETCh:WWQuality
FETCh:WWQuality:PERRor[:MAXimum]?
Function
Queries the maximum phase error measurement.
Query
Range: 0 to 180 degrees, 9.91 E+37
Resolution: 0.1
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:WWQUALITY:PERROR:MAXIMUM?" !Returns maximum phase error.
FETCh:WWQuality:PERRor:ALL?
Function
Queries the phase error measurements.
Results are returned in the following order: minimum, maximum, average, standard deviation
phase error.
Query
Minimum Phase Error
• Range: 0 to 180 degrees, 9.91 E+37
• Resolution: 0.1
Maximum Phase Error
• Range: 0 to 180 degrees, 9.91 E+37
• Resolution: 0.1
Average Phase Error
• Range: 0 to 180 degrees, 9.91 E+37
• Resolution: 0.1
Standard Deviation Phase Error
• Range: 0 to 180 degrees, 9.91 E+37
• Resolution: 0.01
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:WWQUALITY:PERROR:ALL?"
!Returns mininum, maximum, average, and standard deviation phase error.
370
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_fetch_wwquality.fm