W-CDMA Measurements
The following step-by-step procedures explain how to perform specific measurements available on the
W-CDMA Test Application.
The test set and the UE must be sychronized before the UE can generate an uplink signal that can be
measured in the following procedures. See
Detailed descriptions for each measurement and the FDD Test operating mode are provided in the Reference
information on the documentation CD-ROM and at the Agilent 8960 support website
(www.agilent.com/find/8960support).
Setup parameters for each measurement are assumed to be set to their default (full preset) values unless
otherwise stated.
The following measurement procedures are provided:
"Measuring Adjacent Channel Leakage Ratio" on page 555
"Measuring Channel Power" on page 556
"Measuring Loopback Bit Error Ratio" on page 557
"Measuring Thermal Power" on page 558
"Measuring Waveform Quality" on page 559
"Measuring Spectrum Emission" on page 560
"Measuring Occupied Bandwidth" on page 561
"Measuring Code Domain Power and Error" on page 562
"Making Measurements on a UE"
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_man_measurements.fm
W-CDMA Measurements
.
554