DISPlay
DISPlay:BRIGhtness
Function
Sets/queries the test set's display brightness.
A display backlight dimming feature lowers the display brightness after approximately 10
minutes without any manual user interaction with the test set. See
on page
Setting
Range: MEDium | HIGH
Query
Range: MED | HIGH
Factory
HIGH (this parameter is not affected by any reset operation and can only be changed by direct
setting
user access)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"DISPLAY:BRIGHTNESS MEDIUM" !Sets display brightness to medium.
DISPlay:MESSage:MASKable:STATe
Function
Blocks maskable messages from appearing on the test set display display screen but not from the
Message Log. Maskable messages are reported to the Message Log in either state. See
Messages" on page
Setting
Range: On | Off
Query
Range: On | Off
Factory
On
setting
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"DISPLAY:MESSAGE:MASKABLE:STATE OFF"
!Prevents certain messages from appearing on the display.
DISPlay:MODE
Function
Sets/queries the test set's display mode.
"Display Mode (Track/Fast)"
See
Setting
Range: FAST | TRACK
Query
Range: FAST | TRAC
*RST setting
TRACK
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"DISPLAY:MODE FAST" !Sets display mode to fast.
312
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_display.fm
310.
588.
for a description of fast and track display modes.
"Display Backlight Dimming"
"Error