READ
READ:WCPower[:ALL]?
Function
Initiates and fetches a channel power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average channel power.
If the multi-measurement state is on, the result represents the average of a number of channel
power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
• Range: -99 to 99 dBm
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.01 and above
READ:WOBWidth[:ALL]?
Function
Initiates and fetches an occupied bandwidth measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average occupied bandwidth.
If the multi-measurement state is on, the result represents the average of a number of channel
power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
• Range: 0.0 to 10.0 MHz
• Resolution:.0001
Requirements
Test Application Revision: A.02 and above