Test System Synchronization Overview
six.
Many other status bits are provided in the GPIB status subsystem that are useful for synchronization. See
"STATus Subsystem Description" on page
"SYSTem:SYNChronized" on page 532
This specialized command puts a 1 in the test set's output queue, the test set responds to the query by
sending a 1 to the external controller indicating that all prior sequential commands have completed, and all
prior overlapped commands have at least begun execution. The condition bit is set then cleared. See
"STATus:OPERation Register Bit Assignments" on page
and six.
"*OPC"
and
"*OPC?" on page
Note: These commands look at all of the test set's operations collectively. Because multiple processes are
likely to be executing at the same time, it is recommended that the other commands above be used instead.

Related Topics

"Call Processing State Synchronization" on page 1
"Measurement Event Synchronization" on page 201
"Call Processing Event Synchronization" on page 1
"SYSTem:SYNChronized" on page 532
200
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_prog_sync_gen.fm
442.
537, and
"*WAI" on page 538
451. This command supports synchronization four
(not recommended)