CALL:CONNected
CALL:CONNected:DPCHannel[:SLEVel]
Function
Sets/queries the power offset of the dedicated physical channel to use while a call is connected.
This command sets the power offset level and turns on the power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this value.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-3.30 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CONNECTED:DPCHANNEL:SLEVEL -7" !Sets DPCH connected power offset.
CALL:CONNected:DPCHannel:LEVel
Function
Sets/queries the dedicated physical channel's power offset to use while a call is connected. This
command sets the power offset level and turns on the power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this value.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-3.30 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CONNECTED:DPCHANNEL:LEVEL -7"