CALL:CCPChannel
CALL:CCPChannel:PRIMary:STATe
Function
Sets/queries the primary common control physical channel's state.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this setting.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
1 (on)
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CCPCHANNEL:PRIMARY:STATE ON"
CALL:CCPChannel:SECondary[:SLEVel]
Function
Sets/queries the initial power offset of the secondary common control physical channel . This
command sets the power offset level and turns on the power offset.
This setting cannot be changed while in active cell operating mode. Attempting to change this
setting while in this mode results in the following error message: "Command Rejected. Change
Not Allowed in Active Cell Mode." Use the cell off operating mode when changing this setting.
Setting
Range: -20.00 dB to 0 dB
Resolution: 0.01
Query
Range: -20.00 dB to 0 dB
Resolution: 0.01
*RST Setting
-6.9 dB
Requirements
Test Application Revision: A.02 and above.
Programming Example
OUTPUT 714;"CALL:CCPCHANNEL:SECONDARY:SLEVEL -10DB" !Sets P-CCPCH initial power offset to -10 dB.