Agilent Technologies 8960 Series Reference Manual page 338

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

FETCh:WCPower
FETCh:WCPower[:ALL]?
Function
Queries a channel power measurement result.
This query returns an integrity indicator followed by the average channel power measurement
result.
If the multi-measurement state is on, the result represents the average of a number of channel
power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
• Range: -99 to 99 dBm, 9.91 E+37
• Resolution: 0.01 dBm
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:ALL?" !Returns integrity, channel power measurement.
FETCh:WCPower:ICOunt?
Function
Queries the intermediate count of measurements completed in a multi-measurement.
To set up multi-measurements, see
Query
Range: 0 to 999
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:ICOUNT?"
!Returns the intermediate count of channel power multi-measurements.
FETCh:WCPower:INTegrity?
Function
Queries the measurement integrity indicator for the last channel power measurement completed.
Zero indicates a normal measurement. See
about non-zero integrity results.
Query
Range: 0 through 23
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:INTegrity?"
!Queries the integrity indicator for the channel channel power results.
338
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_fetch_wcpower.fm
"SETup:WCPower:COUNt[:SNUMber]"
"SETup:WCPower:COUNt[:SNUMber]"
"Integrity Indicator" on page 190
.
.
for information