CALL:MS:POWer

:MS
:POWer
CALL
"Diagram Conventions" on page 210
CALL:MS:POWer:TARGet[:AMPLitude]?
Function
Sets/queries the UE (mobile station) target transmitter power.
When operating in FDD test mode, this setting sets the UE expected power into the test set when
using automatic receiver control. This setting does not control the UE output power in any way.
The user must set this value when making measurements that do not provide autoranging to
prevent an over range or an under range measurement situation. See
Mode"
When operating in Active Cell mode, if the uplink closed loop power control (UL CL Power Ctrl)
is in the active bits mode, the instrument will measure the mobile's transmit power and it will
send the required transmit power control bit patterns to the mobile. This setting does not directly
set the transmit power of the mobile, but it will drive the mobile power to the setting if closed
loop power is active. See
As in FDD Test operating mode, this setting has the secondary function of setting the test set
receiver's expected power level when the receiver power control state is set to AUTO.
Query
Range: -50 dBm to +24 dBm
Resolution: 1 dB
*RST Setting
24 dBm
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"CALL:MS:POWER:TARGET:AMPLITUDE -20"
:TARGet
[:AMPLitude]
.
"CALL[:CELL]:CLPControl:UPLink:MODE"
<sp><num value>[DBM]
?(returns num value)