Agilent Technologies 8960 Series Reference Manual page 298

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

CALL:STATus
CALL:STATus:CELL:POWer[:AMPLitude]?
Function
Queries the current cell power level.
Cell power is set using the
Query
Range:
• -170dBm/3.84 MHz to +37dBm/3.84 MHz, 9.91 E+37
(-120 dBm/3.84 MHz to -13 dBm/3.84 MHz actual hardware range with amplitude offset = 0)
The actual power range is defined by adding the value of the associated amplitude offset to
the range above. For example, if amplitude offset is -3.5 dB, then the power range will be
-123.5 dBm/3.84 MHz to -16.5 dBm/3.84 MHz.
Resolution: 0.01 dBm/3.84 MHz
*RST Setting
-115.00 dBm
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"CALL:STATUS:CELL:POWER:AMPLITUDE?"
CALL:STATus:CELL:POWer:STATe?
Function
Queries the current state of the cell power setting.
Cell power state is set using the
Query
Range: 1|0
*RST Setting
1 (on)
Requirements
Test Application Revision: A.01 and above
Programming Example
OUTPUT 714;"CALL:STATUS:CELL:POWER:STATE
298
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_call_status.fm
"CALL[:CELL]:POWer"
"CALL[:CELL]:POWer"
?"
commands.
commands.