SETup:WCDomain
SETup:WCDomain:TIMeout:TIME
Function
Sets/queries code domain measurement's timeout setting to be used when the timeout state is set
to ON (see
Setting
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query
Range: 0.1 to 999.9
*RST Setting
20 seconds
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCDOMAIN:TIMEOUT:TIME 5 S"
SETup:WCDomain:TSLot
Function
Sets/queries the timeslot to test in the code domain test.
Setting
Range: 0 to 14
Query
Range: 0 to 14
*RST Setting
0
Requirements
Test Application Revision: A.02 and above
Programming Example
OUTPUT 714;"SETUP:WCDOMAIN:TSLot"
420
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_hpib_setup_wcdomain.fm