4. Hygienic preparation
4.1 Cleaning and disinfecting after use
Caution!
This point is described in Chapter 5 "Cleaning and Disinfection" of the SOMNOsmart 2, SOMNOset or
SOMNOsoft + operating instructions.
Hygienic preparation of the device after repairs and for a new patient is described below.
4.2 Cleaning and disinfection after repair
In the event of repairs the following work should be performed by the qualified dealer.
Caution!
Be sure to follow the disinfectant manufacturer's
instructions (8.3, page 45). You are
recommended to wear suitable gloves (e.g.
household or disposable gloves) during
disinfection procedures.
•
Wipe outside of device and power cord with
TERRALIN disinfectant.
4.3 Cleaning and disinfecting on change of patient
When making the device hygienic for a new patient, the following steps must be carried out:
Caution!
Be sure to follow the disinfectant manufacturer's
instructions (8.3, page 45). You are
recommended to wear suitable gloves (e.g.
household or disposable gloves) during
disinfection procedures.
•
Wipe outside of device and power cord with
TERRALIN disinfectant. Dispose of hose and
mask system and carrying bag, and replace
with new parts.
•
Open device as described in Chapter 7.2.
•
Clean the inside of the device and filter
housing with a vacuum cleaner. Thoroughly
clean any specially dirty parts.
•
Clean hoses, cap and mask as described in the
operating instructions or replace with new parts
(depending on condition).
•
Open device as described in 7.2.
•
Replace the coarse dust filter and fine filter
+
as described in Chapter 3.2.
46
•
Clean the inside of the device and filter
housing with a vacuum cleaner. Thoroughly
clean any specially dirty parts.
•
Close device as described in 7.3.
•
Replace the coarse dust filter and fine filter
+
as described in Chapter 3.2.
46
•
Spray-disinfect device outlet
and filter cover
29 twice
LIQUID", taking care to wait for the prescribed
disinfection time in each case. At the
beginning of the disinfection time, also wipe
the accessible surfaces with a cloth moistened
with "MICROZID LIQUID".
Caution!
No disinfectant residues must be left in the
pressure measurement stub of the device outlet
port
; if necessary the pressure measurement
7
stub must be blown dry.
•
Replace OPS delivery tube
SOMNOsmart 2 and SOMNOset).
Hygienic preparation
45
45
, filter holder
7
44
with "MIKROZID
(only
61
13