Hygienic Preparation; Intervals - Weinmann SOMNOsmart 2 Description And Instructions For Use

Table of Contents

Advertisement

Available languages

Available languages

5. Hygienic preparation

5.1 Intervals

This product may contain disposable items.
items are intended to be used only once. So use these
items only once and do
ing disposable items may impair the functionality and
safety of the product and lead to unforeseeable reac-
tions as a result of ageing, embrittlement, wear, ther-
mal load, the effects of chemical processes, etc.
The mask
14
with thorough cleaning being carried out accord-
ing to the operating instructions once a week.
Check the filters regularly:
clean / change the coarse dust filter once a
week;
change the fine filter depending on the level
of soiling, but at least every 250 hours (filter
change indicator
Regularly wipe clean the housing of the
SOMNOsmart 2 and the filter compartment cover
.
28
The hose system
to the level of soiling, however at least once a
month. The pressure measurement tube
changed at least every 6 months and earlier in the
case of soiling (see "8.3 Change pressure mea-
surement tube" on page 38).
The exhalation system
according to the operating instructions after each
use.
Wash the headgear
necessary.
reprocess them. Reprocess-
not
should be cleaned daily after usage,
).
18
should be cleaned according
12
should be cleaned
13
and/or headbands as
15
Hygienic preparation
Disposable
must be
11
27
EN